Model: YTAT-101 | YTAT-216
Main Parameters:
Temperature range: RT+25℃~200℃(300℃)
Temperature fluctuation: ±0.3C
Temperature resolution: 0.1℃
Air ventilation volume: 5-200 (3-250) time/H
Heating time: 45 min
N=3590(X-Y)x2/V · ∆ f · D
Digital display or program touch screen controller
The ventilation chamber is suitable for heat resistance test of electrical insulating materials, electronic spare parts, the gas aging test of plasticized products, to check its adaptability to storage and operation under high-temperature environment conditions, compare the properties after aging in the air under simulated high temperature and atmospheric pressure.
Yuanyao's ventilation chamber meets JB/T7444 "air heat aging test chamber" and GB/T3512-2001 test standard requirements. Welcome to contact us for the ventilation aging machine price.
Model No. | YTAT-101 | YTAT-216 |
Test area size W×H×D(cm) | 45×50×45 | 60×60×60 |
Exterior chamber size W×H×D(cm) | 105×150×60 | 120×160×75 |
Temperature range | Room temperature RT+25℃~200℃(300℃) | |
Temperature fluctuation | ±0.3℃ | |
Temperature uniformity | ±1%℃ | |
Temperature reslution | 0.1℃ | |
Heating-up time | 30min | 40min |
Exterior chamber material | Cold-rolled sheet + powder coating | |
Interior chamber material | SUS#304 stainless steel plate | |
Insulation material | PU+ fiberglass wool | |
Wind circulation | Centrifugal fan air circulation system | |
Heating system | SUS# Stainless steel high-speed heater | |
Controlling system | Button type digital display controller | |
Air exchange rate | 5~200(3~250)times/hour | |
Other components | Control status indicator, turntable/compartment shelf (2 freely adjustable) | |
Safety protection device | Phase sequence protection, over-temperature protection, overload protection | |
Power supply | AC 3Ψ380V 60/50Hz | |
Customization service | Non-standard or special requirements can be customized. |
Intelligent program microcomputer controller with timing function, temperature control is accurate and reliable.
Equipped with tempered glass window for observation of the conditions inside the chamber. at any time.
Equipped with low-speed rotary table.
Insulation is made of ultra-fine glass fiber insulation cotton to avoid excessive surface temperature of the casing.